This page is to inform owners of ZEISS Ophthalmic Diagnostics Instruments about:
Microsoft Windows Print Spooler Vulnerability CVE-2021-34527 (“PrintNightmare“) Cybersecurity Update
PrintNightmare is a vulnerability affecting Microsoft Windows operating systems (OS). As stated on Microsoft’s web site : “A remote code execution vulnerability exists when the Windows Print Spooler service improperly performs privileged file operations. An attacker who successfully exploited this vulnerability could run arbitrary code with SYSTEM privileges. An attacker could then install programs; view, change, or delete data; or create new accounts with full user rights.“
PrintNightmare does not affect safety and performance on any of the ZEISS devices. However, ZEISS recommends updating devices with the Microsoft patch and/or registry settings, as applicable, to ensure continued cybersecurity.
ZEISS has analyzed the impact of the vulnerability on ZEISS products running Windows OS and only devices listed below must run the update.
Refer to the table below for the respective update instructions to close this vulnerability for your corresponding ZEISS device. Please contact your local service team if you need support updating your device.
-
ModelOSLink to Instructions for Updating ZEISS DeviceLink to Download (use Internet Explorer or Edge)ZEISS CIRRUS
ZEISS CIRRUS 500/5000/6000
Windows 10ZEISS CIRRUS 400/4000/500/5000Windows 7
Follow instructions (see column to the left titled “Link to Instructions for Updating ZEISS Device”) -
ModelOSLink to Instructions for Updating ZEISS DeviceLink to Download (use Internet Explorer or Edge)ZEISS CIRRUS photo 600/800ZEISS CIRRUS photo 600/800Windows 10ZEISS CIRRUS photo 600/800Windows 7Follow instructions (see column to the left titled “Link to Instructions for Updating ZEISS Device”)
-
ModelOSLink to Instructions for Updating ZEISS DeviceLink to Download (use Internet Explorer or Edge)ZEISS ATLAS 9000
ZEISS ATLAS 9000
Windows 10ZEISS ATLAS 9000Windows 7Follow instructions (see column to the left titled “Link to Instructions for Updating ZEISS Device”) -
ModelOSLink to Instructions for Updating ZEISS DeviceLink to Download (use Internet Explorer or Edge)ZEISS CLARUS 500/700ZEISS CLARUS 500/700Windows 10
-
ModelOSLink to Instructions for Updating ZEISS DeviceLink to Download (use Internet Explorer or Edge)ZEISS HFA3ZEISS HFA3Windows 10
Microsoft Windows Print Spooler Vulnerability CVE-2021-34527 (“PrintNightmare“) Cybersecurity Update For Windows 10 OS
Note: the HFA3 is shipped with the instrument running in kiosk mode. To install the patch, disable Kiosk mode (Refer to instructions in the HFA3 IFU to access the Windows desktop), login as the IT Administrator user, run the “Kiosk OFF” shortcut on the desktop, and reboot the instrument.Windows 7Microsoft Windows Print Spooler Vulnerability CVE-2021-34527 (“PrintNightmare“) Cybersecurity Update For Windows 7 OS
Note: the HFA3 is shipped with the instrument running in kiosk mode. To install the patch, disable Kiosk mode (Refer to instructions in the HFA3 IFU to access the Windows desktop), login as the IT Administrator user, run the “Kiosk OFF” shortcut on the desktop, and reboot the instrument.Follow instructions (see column to the left titled “Link to Instructions for Updating ZEISS Device”) -
ModelOSLink to Instructions for Updating ZEISS DeviceLink to Download (use Internet Explorer or Edge)ZEISS PLEX Elite 9000ZEISS PLEX Elite 9000Windows 10ZEISS PLEX Elite 9000Windows 7Follow instructions (see column to the left titled “Link to Instructions for Updating ZEISS Device”)
-
ModelOSLink to Instructions for Updating ZEISS DeviceLink to Download (use Internet Explorer or Edge)ZEISS PRIMUS 200ZEISS PRIMUS 200Windows 10: serial numbers starting with 200-3XXXX and 200-5XXXXZEISS PRIMUS 200Windows 7: serial numbers starting with 200-0XXXX and 200-2XXXXPlease contact your Service TechnicianN/A
-
ModelOSLink to Instructions for Updating ZEISS DeviceLink to Download (use Internet Explorer or Edge)ZEISS IOLMasterZEISS IOLMaster 5006. Generation, WinPOSZEISS IOLMaster 5.5ZEISS IOLMaster 700
-
ModelLink to Instructions for Updating ZEISS DeviceLink to Download (use Internet Explorer or Edge)ZEISS SL Imaging Solution
ZEISS SL Imaging Solution
3 Files to download (unpack from zip):
- Install_Patch_PrinterNightmare_CVE-2021-34527_ Win-10.cmd
- LGPO.exe
- PrintNightmare_1.lgpo.txt
-
ModelLink to Instructions for Updating ZEISS DeviceLink to Download (use Internet Explorer or Edge)ZEISS SL Workstation
ZEISS SL Workstation
3 Files to download (unpack from zip):
- Install_Patch_PrinterNightmare_CVE-2021-34527_ Win-7.cmd
- LGPO.exe
- PrintNightmare_1.lgpo.txt
-
ModelOSLink to Instructions for Updating ZEISS DeviceLink to Download (use Internet Explorer or Edge)ZEISS VISUCAMZEISS VISUCAM 224ZEISS VISUCAM 524ZEISS Visucam 200
Computer 1969-629
ZEISS VISUCAM PRO NM 2ZEISS VISUCAM NM/FA 2ZEISS VISUCAM 500 -
ModelLink to Instructions for Updating ZEISS DeviceLink to Download (use Internet Explorer or Edge)ZEISS VISUPACZEISS VISUPAC 500
3 Files to download (unpack from zip):
- Install_Patch_PrinterNightmare_CVE-2021-34527_ Win-7.cmd
- LGPO.exe
- PrintNightmare_1.lgpo.txt