Technical data
Laser type:
|
femtosecond laser
|
||
Available treatment options:
|
Flap, SMILE® pro, CIRCLE®, ICR, Keratoplasty |
||
Digital assistant systems:
|
Centration aid CentraLign®
|
||
---|---|---|---|
Cyclotorsion adjustment OcuLign®
|
|||
Nomogram tool VISULYZE®5 |
|||
Optical data
|
|||
Maximum laser repetition frequency:
|
2 MHz
|
||
Wavelength:
|
1043 nm
|
||
Pulse duration:
|
220–580 fs
|
||
Surgical microscope
|
|||
Magnification:
|
0.7x
|
||
Factors for magnification change:
|
0.4 / 0.6 / 1.0 / 1.6 / 2.5
|
||
Eyepiece magnification:
|
12.5x (10x) |
||
Filters:
|
blue, barrier filter (yellow)
|
||
Slit illumination: |
slit width ≤ 0.3 mm / 0.7 mm
|
||
slit height 11.0 mm |
|||
Ambient conditions for operation
|
|||
Temperature:
|
+18 °C to +25 °C
|
||
Humidity:
|
30 % to 70 %
|
||
Measurements
|
|||
Device mass:
|
520 kg
|
||
Minimum floor load:
|
2.5 kN/m2 |
||
Footprint standalone:
|
L x W:1710 mm x 925 mm
|
||
|
|
||